2. Technical specifications
| Pressure connection: |
Measuring connection M16, max. 315 bar (Minimess) |
| Suction connection: |
Screw connection 6 L, max. 10 bar (140 psi) |
| Return line: |
Screw connection 6 L |
| Pressure fluctuation: |
permissible |
| Medium: |
hydraulic fluids, fuels, water precondition of no second phase |
| Viscosity range: |
max. 500 mm², on suction side max. 68 mm²/s |
Temperature range ambience: Temperature range fluid: |
0 to +50 °C 0 to 80 °C |
| Sensor flow rate: |
30 ml/min |
| Flushing volume flow: |
30 ml/min |
| Measuring volume: |
10 to 100 ml, adjustable (10 ml steps) |
| Volume prior to counting: |
10 to 100 ml, adjustable in (10 ml steps) |
| Counting time: |
30 s |
| Cycling time: |
1 to 99 min |
| Sensor: |
Laser diode sensor |
| Tank volume: |
1 l |
| Indication acc. SAE AS 4059: |
>4/>6/>14/>21/>38/>70 µm(c) |
| Measuring range acc. SAE classes: |
000 to 12 |
| Indication acc. ISO 4406: 1999 |
>4/>6/>14 µm(c) |
| Measuring range ISO classes: |
1 to 24 |
| Calibration: |
acc. ISO 11171: 1999 |
| Power supply: |
100 to 230 VAC; 50/60 Hz; accumulator optional |
| Dimensions: |
320x450x300 mm (HxWxD) |
| Weight: |
approx. 12 kg | |
 |
3. Technical manual
The portable particle counter PiC 9300 consists of a sensor with pump/volume regulation unit and tank with overflow safety device. The counted results are displayed on a touch screen monitor and can be printed with the integrated thermal printer. Thanks to the strengthened polyamide plastic moulding material, the housing has an optimal compatibility against all common pressure fluids. The electrical connections are placed at the backside of the unit and are therefore properly protected. For the electrical connecting cable, accumulator and hoses is a separate compartment available. Due to the removable cover is an optimal access to all control elements, a safe storage and transport ensured. For data transfer to a PC is an USB and a serial interface available. Also, it is possible to store the data via a second USB interface in an Excel compatible format. The language can be selected between German, English and French. _
4. Options
Measuring and analysing with "LOG and SHOW" software for data transfer to notebook (graphic illustration). Accumulator version for off-grid power supply. Bottle sampling- and adapter-kit.
|